The particle sensor

The monitor measures particles with a Plantower PMS5003. It pulls air through with a small fan and counts particles with a laser. Both parts wear out with use, so we let the sensor rest between short bursts. This page explains that choice, what averaging can and can’t fix, and the other sensors we looked at. Every claim links to its source, and we’ve noted where a source is weaker.

These come from Plantower’s PMS5003 manual, V2.3 (2016). We keep a copy in the firmware repo under docs/datasheets/, with its checksum and page numbers.

How often it reports Every 2.3 s when the air is steady; every 200–800 ms when it changes fast
Warm-up Readings are stable “at least 30 seconds after the sensor wakeup”
Sleep A UART sleep command (or pulling its SET pin low) stops the fan and laser; a wake command starts them again
Power 100 mA running, 200 µA asleep
Expected life MTTF (mean time to failure) of 3 years, with no test conditions given
Sensor-to-sensor spread ±10 µg/m³ up to 100 µg/m³, ±10 % above that
Sleep cycling Plantower’s endurance test toggles sleep every 2 s for 72 hours, so frequent rests are safe

Two honest caveats from the same manual: Plantower only promises that its sensors agree with each other, not with reference instruments, and it rates the sensor for normal indoor air, asking for extra protection outdoors.

A sensor that runs all day ages fast. The best long-term look we found is a large study of PurpleAir monitors, each with two PMS5003s running nonstop (deSouza et al., 2023). The two sensors in a monitor disagree more often as they age, and their bias rises sharply after about 3.5 years, more so in hot, humid places. That lines up with the 3-year MTTF in the datasheet.

Sleep stops both the fan and the laser, so a sensor that runs a fifth of the time spends a fifth as many hours on its wearing parts. Here’s the cycle:

  1. Every 5 minutes, we wake the sensor.
  2. We ignore its readings for 30 s while the fan settles, as the datasheet asks.
  3. We average every reading for the next 30 s, about 13 of them.
  4. It sleeps for the remaining 4 minutes.

Each 10-minute upload averages two bursts, about 26 readings. The fan and laser run 60 s in every 5 minutes, so we expect roughly five times longer life. That’s an estimate from running hours, not a measured result. No study we found has tested it directly.

What we give up: a short spike between bursts, like a burst of cooking smoke, can be missed. For readings shared every 10 minutes and viewed as hourly averages, we think that’s a fair trade.

Averaging steadies readings, but doesn’t correct them

Section titled “Averaging steadies readings, but doesn’t correct them”

Averaging many readings smooths out random noise, so one odd reading can’t swing an upload. It can’t fix a sensor that reads high across the board.

Raw PMS5003 readings run about 40 % high in most of the US, and humidity makes it worse, because damp particles swell and look bigger to the laser. The US EPA built a correction from co-located reference monitors that brings the typical error from 8 down to 3 µg/m³ on 24-hour averages (Barkjohn et al., 2021):

PM2.5 = 0.524 × PM2.5 (CF=1) − 0.0862 × relative humidity + 5.75

We don’t apply this yet. It needs the sensor’s CF=1 values (we send its “atmospheric” values today) and a humidity reading next to the sensor, which the monitor already has. It’s planned as a separate change, so for now please treat readings as useful for trends rather than exact values.

We compared the PMS5003 with five other sensors before deciding to stay with it for now.

Sensor What it offers Why we’re not switching
Sensirion SPS30 Rated for “more than ten years”, resists dust build-up, cleans its own fan weekly, reports every second It lasts longer, but isn’t clearly more accurate: in a field test it also read high, and it needs the same humidity correction. It costs about three times as much and needs a new driver and connector. It’s our strongest candidate for the purpose-built board.
Sensirion SEN66 / SEN5x Particles plus temperature, humidity, VOC and NOx (and CO₂ on the SEN66) in one module We’d pay for gas sensing we don’t use yet, and it overlaps our SHT30. Worth another look if VOC or CO₂ joins the roadmap.
Nova SDS011 Popular with community networks; has a built-in timed sleep mode Its laser is rated “up to 8000 hours”, well short of the PMS5003’s 3-year MTTF (about 26,000 hours), and it’s bigger.
Honeywell HPMA115S0 Compact, UART like ours Its stated PM2.5 accuracy (±15 %) is looser than the PMS5003’s, and Honeywell’s own documents disagree on its life (20,000 hours or 10 years).
Plantower PMSA003I The same Plantower design with an I²C connection Same fan, laser and bias as the PMS5003, so it wouldn’t help with any of this.

A year-long study ran eight SPS30s and eight PMS5003s next to a reference monitor in Southampton (Bulot et al., 2023). Both reached reference-grade PM2.5 only after calibration that included humidity. That’s the main lesson for us: calibration matters more than the choice between these two sensors, which is why it’s our next step.

How much weight each source carries:

What we couldn’t verify, so we don’t rely on it: the “8,000-hour laser” figure often quoted for the PMS5003 (it isn’t in Plantower’s manual and seems to come from the SDS011’s datasheet), and which sensor did better in the Southampton study (we couldn’t access its detailed results).